点击数:66412015-06-04 18:02:53 来源: PACS系统-超声工作站网络版-内窥镜工作站-放谢工作站-DICOM工作站-pacs系统-超声医学影像工作站-超声影像工作站-超声工作站软件-精子分析系统-泰融科技
缺陷描述 | 新换的电脑,新装的纯系统,用到第三天,出现打印黑片的现像。 |
解决方法 |
看log发现,西门子的设备,dicom协议与常规设备有区别,就是打印开始到打印结束后一直在握手,没有断开连接,从log中能看到,间隔大约30秒就发送一次握手测试,这样,资源占用太大,得不到释放,会造成死机、打印黑胶片等。 解决方法:联系西门子工程师,修改设备的注册表中的一个项目,重新启动设备即可(必须的!!!)。修改后的log文件中,会间隔5分多钟执行一次握手测试,log中会提示【释放资源】,这样就可以了。 |
昵称:游客 curry 9 | IP:27.153.*.*. | 2022-10-24 11:30:52发表 |
I and also my pals appeared to be viewing the great thoughts found on the blog and immediately I got a terrible suspicion I had not expressed respect to the blog owner for those strategies. The ladies are actually certainly joyful to see all of them and have now definitely been loving those things. Many thanks for really being really accommodating and for deciding upon these kinds of important tips most people are really desirous to understand about. Our honest regret for not expressing gratitude to sooner. | ||